user-circle-o search facebook linkedin2 twitter document-download info
Exterior view of the American Biotech Supply PH-ABT-NSF-UCFS-0504G undercounter medical freezer, featuring a compact white cabinet with a glass door, digital temperature display, and interior shelving, designed for pharmaceutical and clinical cold storage. NSF/ANSI 456 vaccine refrigerator 5.2 cu. ft. glass door
Exterior view of the American Biotech Supply PH-ABT-NSF-UCFS-0504G undercounter medical freezer, featuring a compact white cabinet with a glass door, digital temperature display, and interior shelving, designed for pharmaceutical and clinical cold storage. NSF/ANSI 456 vaccine refrigerator 5.2 cu. ft. glass door

5.2 Cu. Ft. Glass Door Freestanding Vaccine Refrigerator Certified to NSF/ANSI 456

Product Number: PH-ABT-NSF-UCFS-0504G
ETL_Sanitation-listed3intertek-etl-listed-us-mark-seeklogo31 (002)*ul*Green-Leaf-ABS

These premier undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

These glass door freestanding refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

*-one or more of these certifications may apply to this unit.

 

Description Single Glass Door Pharmacy/Vaccine Undercounter Refrigerator Freestanding
Operational environment Indoor use only. Optimal operating range: +18°C to +26°C (+65°F to +78°F), 70% RH
Storage capacity 5.2 cu. ft. gross volume
Door One swing glass door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
Shelves Three shelves (two adjustable/one fixed) with guard rail on back
Mounting and Installation Leveling legs. Note: 4″ of clearance on all sides must be maintained for adequate ventilation
Interior lighting Shielded, switched LED lighting, full coverage, balanced spectrum
Airflow management Forced Air technology, patent pending
External probe access Rear wall port (3/8″) dia.
Insulation Cabinet is foamed-in-place with EPA compliant high density urethane foam
Exterior materials White powder coated steel
Access control Pyxis®, Omnicell® and AcuDose RX® compatible
General warranty Two (2) years parts and labor warranty, excluding display probe calibration
Compressor warranty Five (5) years compressor warranty
Product Weight 96 lbs.
Shipping Weight 132 lbs.
Rated Amperage 1.3 Amps
Power Plug/Power Cord NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
Facility Electrical Requirement 110-120V AC: 15 A (minimum)

Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety. Energy Star Certified

 

Pharmacy refrigerator/freezer toolkit and temperature logs

Agency Listing and Certification
 

Included Accessories

 

Compressor Hermetic, high performance
Refrigerant EPA SNAP compliant, R600a, Isobutane
Condenser Tube and grid construction, fanless
Evaporator Plate wall
Defrost Cycle optimized, zero energy
Uniformity¹ (Cabinet air) +/- 1.4°C
Stability² (Cabinet air) +/- 1.3°C
Maximum temperature variation (Cabinet Air) +/-1.7°C
Temperature rise after 8 sec door openings Temperature did not exceed 6.5°C at any probe for all required NSF/ANSI 456 testing scenariosᶟ
Recovery after 3 min door opening All probes recover to under 8°C within 6 min.
Energy consumption 1.15 KWh/day⁴
Average heat rejection 1.67 KWh/day (237 BTU/h)⁴
Noise pressure level (dBA) 41 or less installed
Pull down time to nominal operating temp 36 min
Controller technology Parametric, microprocessor, LED display with 0.1°C resolution
Temperature setpoint range 1°C to 10°C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
Calibration Calibrated using a NIST traceable device, certificate included
External alarm connection State switching remote alarm contacts
Alarms Visual and audible indicators, High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage

Inquire About This Product