4.6 cu. ft. Built-in Glass Door Vaccine Refrigerator Left Hinged Certified to NSF/ANSI 456

Product Number: PH-ABT-NSF-UCBI-0404G-LH
intertek-etl-listed-us-mark-seeklogo31 (002)*ul* Green-Leaf-ABS

These premier built-in undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

The glass door, left hinged refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

*-one or more of these certifications may apply to this unit.

 

Description Single Glass Door Pharmacy/Vaccine Undercounter Refrigerator Built-In
Operational environment Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
Storage capacity 4.6 cu. ft. gross volume
Door One swing glass door, self-closing, left hinged, non-reversible, magnetic sealed gasket, keyed lock
Shelves Three shelves (two adjustable/one fixed) with guard rail on back
Mounting Low profile roller wheels and leveling legs
Interior lighting Shielded, switched LED lighting, full coverage, balanced spectrum
Airflow management Forced Air technology, patent pending
External probe access Rear wall port (1/2″) dia.
Insulation Cabinet is foamed-in-place with EPA compliant high density urethane foam
Exterior materials White powder coated steel
Access control Pyxis®, Omnicell® and AcuDose RX® compatible
General warranty Two (2) years parts and labor warranty, excluding display probe calibration
Compressor warranty Five (5) years compressor warranty
 Exterior Width 23 7/8″
 Exterior Depth 26″
 Exterior Height 33 3/8″
Product Weight 100 lbs.
Shipping Weight 140 lbs.
Rated Amperage 1.74 Amps
Power Plug/Power Cord NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
Facility Electrical Requirement 110-120V AC: 15 A (minimum)
Agency Listing and Certification Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
Included Accessories Pharmacy refrigerator/freezer toolkit and temperature logs

 

Uniformity¹ (Cabinet air) +/- 0.8°C
Stability² (Cabinet air) +/- 1.2°C
Maximum temperature variation (Cabinet air) +/- 1.4°C
Temperature rise after 8 sec door openings Temperature did not exceed 6.4°C at any probe for all required NSF/ANSI 456 testing

protocolsᶟ

Recovery after 3 min door opening All probes recover to under 8°C within 4.8 min.
Energy consumption 1.15 KWh/day⁴
Average heat rejection 1.57 KWh/day (224 BTU/h)⁴
Noise pressure level (dBA) 43 or less installed
Pull down time to nominal operating temp 35 min
Controller technology Parametric, microprocessor, LED display with 0.1°C resolution
 Temperature setpoint range 1°C to 10°C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
Calibration Calibrated using a NIST traceable device, certificate included
External alarm connection State switching remote alarm contacts
Simulator ballast Glass bead thermal media

 