PH-ABT-NSF-UCFS-0104G Ext Image PH-ABT-NSF-UCFS-0104G Int Image
PH-ABT-NSF-UCFS-0104G Ext Image
PH-ABT-NSF-UCFS-0104G Int Image

1 cu. ft. Glass Door Freestanding Countertop Vaccine Refrigerator Certified to NSF/ANSI 456

Product Number: PH-ABT-NSF-UCFS-0104G
EnergyStarETL_Sanitation-listed3intertek-etl-listed-us-mark-seeklogo31 (002)*ul*Green-Leaf-ABS

These premier countertop refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

These glass door freestanding refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

*-one or more of these certifications may apply to this unit.

Description Single Glass Door Pharmacy/Vaccine Countertop Refrigerator Freestanding
Operational environment Indoor use only. Optimal operating range: +18°C to +26°C (+65°F to +78°F), <70% RH
Storage capacity 1 cu. ft. gross volume
Door One swing glass door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
Shelves Three shelves (two adjustable/one fixed) with guard rail on back
Mounting and Installation Leveling legs. Note: 4" of clearance on all sides must be maintained for adequate ventilation
Interior lighting Shielded, switched LED lighting, full coverage, balanced spectrum
Airflow management Forced Air technology, patent pending
External probe access Rear wall port (23/32") dia.
Insulation Cabinet is foamed-in-place with EPA compliant high density urethane foam
Exterior materials White powder coated steel
Access control Pyxis®, Omnicell® and AcuDose RX® compatible
General warranty Two (2) years parts and labor warranty, excluding display probe calibration
Compressor warranty Five (5) years compressor warranty
 Exterior Width 17 1/4" 
 Exterior Depth 19 1/4" 
 Exterior Height 21 1/4" 
Product Weight 66 lbs.
Shipping Weight 94 lbs.
Rated Amperage 0.9 Amps
Power Plug/Power Cord NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
Facility Electrical Requirement 110-120V AC: 15 A (minimum)
Agency Listing and Certification Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
Included Accessories Temperature monitor device (TMD) complies with the current CDC guidelines, with 3 years certification of calibration, “buffered” probe in the product simulated solution, min/max memory. Field installable, and visual & audible temp alarm

Pharmacy refrigerator/freezer toolkit and temperature logs

Refrigeration System
Compressor Hermetic, high performance
Refrigerant EPA SNAP compliant, R600a, Isobutane
Condenser Tube and grid construction, fanless
Evaporator Plate wall
Defrost Cycle optimized, zero energy
Performance
Uniformity¹ (Cabinet air) +/- 1.5°C
Stability² (Cabinet air) +/- 0.7°C
Maximum temperature variation +/-1.8°C
Temperature rise after an after 8 sec door openings Temperature did not exceed 6.8°C at any probe for all required NSF/ANSI 456 testing scenariosᶟ
Recovery after 3 min door opening All probes recover to under 8°C within 5.5 min.
Energy consumption 0.71 KWh/day⁴
Average heat rejection 1.01 KWh/day (237 BTU/h)⁴
Noise pressure level (dBA) 34 or less installed
Pull down time to nominal operating temp 35 min

 

Performance data acquired at 22°C ambient, using NSF/ANSI 456 compliant validation ballast probes, empty chamber, during stabilized steady state operation and a DAQ sampling rate of one measurement every 10 seconds

  1. Uniformity is defined as the maximum variance in temperature across all probes at any point in time over the testing period
  2. Stability is defined as the maximum variance in temperature experienced by any single probe over the testing period
  3. Temperature performance for all loaded and unloaded door opening protocols, all alarm, controller and probe requirements as defined in the NSF/ANSI 456 standard for vaccine storage
  4. Data per Energy Star test results or equivalent testing and calculation.  Heat rejection based on daily averages, not continuous operation.  Performance exceeds Energy Star requirements.
ENERGY STAR CERTIFIED
Controller technology Parametric, microprocessor, LED display with 0.1°C resolution
Temperature setpoint range 1°C to 10°C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 standard for Vaccine Storage requirements)
Display probe Calibrated, stainless steel
External alarm connection State switching remote alarm contacts
Alarms Visual and audible indicators
High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
Simulator ballast Glass bead thermal media