user-circle-o search facebook linkedin2 twitter document-download info
A tall, white, single-door medical refrigerator with a glass front, digital temperature display, and five adjustable wire shelves. The fridge stands on wheels and features Medi-Fridge written vertically on the door. The interior is brightly lit.
A tall, white, single-door medical refrigerator with a glass front, digital temperature display, and five adjustable wire shelves. The fridge stands on wheels and features Medi-Fridge written vertically on the door. The interior is brightly lit.

PH-ABT-NSF-16G 16 Cu. Ft. Premier Certified to NSF/ANSI 456 Pharmacy Glass Door Refrigerator

Product Number: PH-ABT-NSF-16G
intertek-etl-listed-us-mark-seeklogo31 (002)*ul *Green-Leaf-ABS

These premier upright refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

These glass door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

*-one or more of these certifications may apply to this unit.

 

Description Single Glass Door Pharmacy/Vaccine Upright Refrigerator
Operational environment Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
Storage capacity 16 cu. ft. gross volume
Door One swing glass door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
Shelves Six shelves (five adjustable/one fixed) with guard rail on back
Mounting 3 1/2″ Swivel Casters(two locking)
Interior lighting Shielded, switched LED lighting, full coverage, balanced spectrum
Airflow management Forced Air technology, patent pending
External probe access Rear wall port (3/4″) dia.
Insulation Cabinet is foamed-in-place with EPA compliant high density urethane foam
Exterior materials White powder coated steel
Access control Pyxis®, Omnicell® and AcuDose RX® compatible
General warranty Two (2) years parts and labor warranty, excluding display probe calibration
Compressor warranty Five (5) years compressor warranty
Product Weight 249 lbs.
Shipping Weight 289 lbs.
Rated Amperage 3 Amps
Power Plug/Power Cord NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine Storage power cord warning label
Facility Electrical Requirement 110-120V AC: 15 A (minimum)
Agency Listing and Certification Certified with the temperature performance requirements as defined in the NSF/ANSI 456 Standard for Vaccine Storage for all testing scenarios. UL, C-UL, ETL, C-ETL listed and certified to UL471 standard, hydrocarbon refrigerant safety.
Included Accessories Pharmacy refrigerator/freezer toolkit and temperature logs

 

Uniformity¹ (Cabinet air) +/- 0.9°C
Stability² (Cabinet air) +/- 1.0°C
Maximum temperature variation (Cabinet air) +/-1.2°C
Temperature rise after 8 sec door openings Temperature did not exceed 4.3°C at any probe for all required NSF/ANSI 456 testing protocolsᶟ
Recovery after 3 min door opening All probes recover to under 8°C within 6.5 min.
Energy consumption 1.25 KWh/day⁴
Average heat rejection 1.97 KWh/day (280BTU/h)⁴
Noise pressure level (dBA) 48 or less installed
Pull down time to 4°C nominal operating temp 30 min

 

Controller technology Parametric, microprocessor, LED display with 0.1°C resolution
Temperature setpoint range 1°C to 10°C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
Calibration Calibrated using a NIST traceable device, certificate included
External alarm connection State switching remote alarm contacts
Alarms Visual and audible indicators, High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
Simulator ballast Glass bead thermal media

 