user-circle-o search facebook linkedin2 twitter document-download info

23 Cu. Ft. Capacity Standard Certified to NSF/ANSI 456 Pharmacy Solid Door Refrigerator

Product Number: PH-ABT-NSF-S23S
intertek-etl-listed-us-mark-seeklogo31 (002)*ul*Green-Leaf-ABS

These standard upright refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

These solid door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

*-one or more of these certifications may apply to this unit.

 

Description Single Solid Door Pharmacy/Vaccine Upright Refrigerator
Operational environment Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
Storage capacity 23 cu. ft. gross volume
Door One swing glass door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
Shelves Five shelves (four adjustable/one fixed) with guard rail on back
Mounting 3 1/2″ Swivel Castors(two locking)
Interior lighting Shielded, switched LED lighting, full coverage, balanced spectrum
Airflow management Forced Air technology, patent pending
External probe access Rear wall port (3/4″) dia.
Insulation Cabinet is foamed-in-place with EPA compliant high density urethane foam
Exterior materials White powder coated steel
Access control Pyxis®, Omnicell® and AcuDose RX® compatible
General warranty One (1) year parts and labor warranty, excluding display probe calibration
Compressor warranty Five (5) years compressor warranty
 Exterior Width 26 7/8″
 Exterior Depth 34 7/8″
 Exterior Width 81 3/4″
Product Weight 216 lbs
Shipping Weight 256 lbs
Rated Amperage 3 Amps
Power Plug/Power Cord NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine Storage power cord warning label
Facility Electrical Requirement 110-120V AC: 15 A (minimum)
Agency Listing and Certification Certified with the temperature performance requirements as defined in the NSF/ANSI 456 Standard for Vaccine Storage for all testing scenarios. UL, C-UL, ETL, C-ETL listed and certified to UL471 standard, hydrocarbon refrigerant safety.
Included Accessories Temperature monitor device (TMD) complies with the current CDC guidelines, with 3 years certification of calibration, “buffered” probe in the product simulated solution, min/max memory. Field installable, and visual & audible temp alarm

Pharmacy refrigerator/freezer toolkit and temperature logs

 

Refrigeration System
Compressor Hermetic, high performance
Refrigerant EPA SNAP compliant, R290, propane
Condenser Fin and tube design, high efficiency fan
Evaporator Fin and tube design, high efficiency fan
Defrost Cycle optimized, zero energy
Performance
Uniformity¹ (Cabinet air) +/- 1.0°C
Stability² (Cabinet air) +/- 1.1°C
Maximum temperature variation +/-1.4°C
Temperature rise after an after 8 sec door openings Temperature did not exceed 6.7°C at any probe for all required NSF/ANSI 456 testing protocolsᶟ
Recovery after 3 min door opening All probes recover to under 8°C within 6.5 min.
Energy consumption 1.32 KWh/day⁴
Average heat rejection 2.21 KWh/day (315 BTU/h)⁴
Noise pressure level (dBA) 49 or less installed
Pull down time to 4°C nominal operating temp 30 min

 

Performance data acquired at 22°C ambient, using NSF/ANSI 456 compliant validation ballast probes, empty chamber, during stabilized steady state operation and a DAQ sampling rate of one measurement every 10 seconds

  1. – Uniformity is defined as the maximum variance in temperature across all probes at any point in time over the testing period
  2. – Stability is defined as the maximum variance in temperature experienced by any single probe over the testing period
  3. – Temperature performance for all loaded and unloaded door opening protocols, all alarm, controller and probe requirements as defined in the NSF/ANSI 456 standard for vaccine storage
  4. – Data per Energy Star test results or equivalent testing and calculation.  Heat rejection based on daily averages, not continuous operation.  Performance exceeds Energy Star requirements.

 

 

Controller technology Parametric, microprocessor, LED display with 0.1°C resolution
Temperature setpoint range 1°C to 10°C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
Display probe Calibrated, stainless steel
External alarm connection State switching remote alarm contacts
Alarms Visual and audible indicators
High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
Simulator ballast Glass bead thermal media