user-circle-o search facebook linkedin2 twitter document-download info
Standard Series
A tall, white upright freezer with a vertical handle on the left side. It has yellow and red warning labels on the front door and a digital temperature display at the bottom. The freezer is from American BioTech Supply. A white, upright refrigerator with its door open, revealing four empty wire shelves inside. The door has a vertical handle and built-in shelf grooves. A small digital temperature display is located at the bottom front panel of the fridge.
A tall, white upright freezer with a vertical handle on the left side. It has yellow and red warning labels on the front door and a digital temperature display at the bottom. The freezer is from American BioTech Supply.
A white, upright refrigerator with its door open, revealing four empty wire shelves inside. The door has a vertical handle and built-in shelf grooves. A small digital temperature display is located at the bottom front panel of the fridge.

Enter your email to receive this document

This field is for validation purposes and should be left unchanged.
Name(Required)
Email(Required)
By submitting a request, you agree and consent to the use of your information in accordance with our Privacy Policy


You maybe required to confirm your email address within 2hrs to complete this request. Check your inbox.

30 cu. ft. Standard Hazardous Location Freezer

Product Number: ABT-HC-EFS-30 ★ NEW PRODUCT
ALSO AVAILABLE IN THESE SIZES4 14 17 20 CU. FT.

This hazardous location freezer meets NFPA and OSHA guidelines for the safe storage of materials in potentially hazardous work environments. This model is the ideal choice for industrial environments requiring strict safety protocols. It delivers 30 cubic feet of spacious storage and utilizes EPA SNAP compliant natural refrigerants for environmentally responsible operation. Powered by microprocessor temperature technology, this model delivers precise and stable temperature control. Please note that this unit must be hardwired for safety. With low energy consumption and high safety standards, this model is a safe and efficient solution for hazardous laboratory locations and industrial settings.

Request a Custom Quote

Get pricing and availability for this product

Gelu

Ask Gelu about the ABT-HC-EFS-30

Here are common questions and answers about this product

Gelu is on every product pageJust ask a question, anytime
What temperature range does this freezer operate in?

The adjustable temperature range is -15° C to -25° C.

Temp Controller tab
How much clearance is required around the unit?

The equipment requires 4" of clearance around the back and sides of the unit for proper ventilation. Improper installation will void the warranty and will lead to temperature maintenance issues, overheating and possible failure.

Temp Controller tab
What are the electrical requirements and does it come with a power cord?

The facility must provide 110–120V AC, a 15A breaker, and a NEMA 5-15 receptacle, and the unit draws 3.3 Amps. A power cord is not included; final power connections must be hardwired to meet local codes.

Description tab
What refrigerant does this freezer use and what type of defrost does it have?

It uses EPA SNAP compliant R290 propane refrigerant and is manual defrost. Defrosting of the unit will need to be performed periodically to prevent severe ice build-up.

Performance tab
What alarms are included on this freezer?

The unit has audible and visual high/low-temperature alarms, a door ajar alarm, and remote alarm contacts for external alarm connection.

Temp Controller tab
What hazardous location certifications does this freezer carry?

It is UL listed per UL 60335-1 and UL 60335-2-89 standards and cUL listed per CSA C22.2. It meets NFPA and OSHA guidelines for hazardous locations, Class 1 Division II, Groups C & D.

Description tab
Answers come from this product's record and datasheet. Settings and procedures come from the owner's manual.

Product details for the ABT-HC-EFS-30

The description, performance figures and controller details, straight from the datasheet.

Need the full datasheet? Use View / Print Product Information above.
Storage capacity (cu. ft)
30 Cu. Ft.
Door
One (1) solid swing, right hinged; stainless steel door handle; magnetic door gasket
Shelves
Four (4) fixed evaporator shelves
Mounting and Installation
Leveling legs
External probe access
3/4″ probe access port on rear wall
Insulation
EPA compliant high density urethane foam
Exterior materials
Powder coated steel
Access control
Keyed door lock
General warranty
Two (2) year parts and labor warranty
Compressor warranty
Five (5) year compressor parts warranty
Product Weight (lbs)
300
Shipping Weight (lbs)
330
Rated Amperage
3.3 Amps
Power Plug/Power Cord
Power cord not included, final power connections must be hardwired to meet local codes
Facility Electrical Requirement
110 – 120V AC, 15A breaker, NEMA 5-15 receptacle
Agency Listing and Certification
UL listed per UL 60335-1 and UL 60335-2-89 standards; c UL listed per CSA C22.2 Meet NFPA and OSHA guidelines for hazardous locations, Class 1 Division II, Groups C & D
Uniformity (Cabinet air)
+/- 3°C
Stability (Cabinet air)
+/- 4°C
Maximum temperature variation (Cabinet air)
8°C Total
Energy Consumption (kWh/day)
5.0 kWh/day
Average Heat Rejection (BTU/hr)
705 BTU/hr
Pull down time to nominal operating temp
70 min (empty)
Compressor
Hermetic, variable speed
Refrigerant
EPA SNAP compliant R290 propane
Condenser
Static – exterior walls of cabinet with auxiliary fan
Evaporator
Evaporator shelves
Defrost
Manual defrost
Controller technology
Microprocessor temperature controller with variable frequency compressor control
Battery Backup
N/A
Display technology
Digital temperature display, sample simulated digital temperature display; .1° resolution
Adjustable Temperature Range
-15° C to -25° C
External alarm connection
Remote alarm contacts
Alarms
Audible and visual high/low-temperature alarms, door ajar alarm, remote alarm contacts
Display Probe
Product simulator bottle with glass bead media
Disclaimers
Defrosting of the unit will need to be performed periodically to prevent severe ice build-up. Indoor use only, +18° C to +26° C (+65° F to +78° F), <70% RH Specifications are subject to change. Equipment requires 4” of clearance around back and sides of the unit for proper ventilation – improper installation will void the warranty and will lead to temperature maintenance issues, overheating and possible failure. Performance claims are based on testing at 22°C ambient using unbuffered air temperatures in an empty freezer at steady state operation. Performance may vary with environmental conditions, product load, or door openings.

Inquire About This Product

This field is for validation purposes and should be left unchanged.
Name(Required)
Email(Required)
Which State is Your Business Located(Required)
US Based Phone Number Only (###) ###-####
Preferred Contact Method(Required)
If you have any additional questions, enter them here
By submitting a request, you agree and consent to the use of your information in accordance with our Privacy Policy
🔥